| Method | Drying with far infrared energy and weighing with an electronic balance |
| Component units | Balance, keypad, and heater |
| Balance specifications | Capacity: Dual range, 0 to 60g and 0 to 340g |
| Readability: 0.001g (0 to 60g) and 0.01g (0 to 340g) |
| Sample pan: Stainless steel, 90mm diameter |
| Moisture analysis specifications | Readability: 0.01% |
| Measuring range: 0 to 100% moisture or solids |
| Measuring accuracy: ±0.1% (May vary depending on sample and condition) |
| Display: Sample weight and percent moisture or percent moisture and time |
| Digital output | Optional EP-50 Electronic Printer prints analysis results and can be programmed to print moisture content at user set time intervals. (EP-50's statistics function cannot be used with EB-340MOC) |
| RS-232C interface via optional IFB-101A interface board, Bi-directional computer link |
| Keypad | Numerical keys and nine function keys to set analysis parameters and display features |
| Heater unit | Heater: Two 150W ceramic heaters |
| Heater area: 100 x 100mm |
| Temperature range of heater surface: Ambient +50 to 500¡C |
| Dimensions and weight | Balance unit: 208W x 325D x 90H mm, 4.9kg |
| Heater unit: 200W x 310D x 155H mm, 4.0kg |
| Keypad: 120W x 110D x 20H mm, 200g |
| Power requirements | 100, 117, 220, 240 V AC ±10% (Selectable), 50/60Hz |
| Power consumption: Balance 8VA, Heater 400VA (average approx.) 1000VA at start of heating cycle |